At a Glance |
|
NIRQuest512 |
NIRQuest512-2.2 |
NIRQuest256-2.1 |
NIRQuest256-2.5 |
Detector: |
Hamamatsu G9204-512 InGaAs linear array |
Hamamatsu G9204-512W InGaAs linear array |
Hamamatsu G9206-256 InGaAs linear array |
Hamamatsu G9208-256 InGaAs linear array |
Detector Range: |
850-1700 nm |
900-2200 nm |
900-2100 nm |
900-2550 nm |
Useable range1: |
900-1700 nm |
900-2200 nm |
900-2050 nm |
900-2500 nm |
Pixels: |
512 |
512 |
256 |
256 |
Pixel size: |
25 m x 500 m |
25 m x 250 m |
50 m x 500 m |
50 m x 500 m |
Entrance aperture: |
25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit)
|
25 m wide slit (standard); also, custom options (10, 50, 100 or 200 m wide slits) or fiber (no slit)
|
25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit) |
25 m wide slit (standard); also, custom options (50, 100 or 200 m wide slits) or fiber (no slit) |
Fiber optic connector: |
SMA 905 |
SMA 905 |
SMA 905 |
SMA 905 |
Grating options (standard): |
Grating NIR3, 150 l/mm, 900-1700 nm |
Grating NIR2, 100 l/mm, 900-2050 nm |
Grating NIR2, 100 l/mm, 900-2050 nm |
Grating NIR1, 150 l/mm, 1075-2500 nm |
Grating options (custom): |
Gratings NIR10, NIR11, NIR12, NIR13 and NIR14 |
Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 |
Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 |
Gratings NIR2, NIR3, NIR10, NIR11, NIR12 and NIR13 |
Longpass filter2: |
OF1-RG830 longpass NIR filter (optional) |
OF1-RG830 longpass NIR filter (optional) |
OF1-RG830 longpass NIR filter (optional) |
OF1-RG830 longpass NIR filter (optional) |
Wavelength range: |
900-1700 nm
w/Grating NIR3 |
900-2150 nm
w/Grating NIR2 |
900-2050 nm
w/Grating NIR2 |
900-2500 nm
w/Grating NIR1 |
Optical resolution (FWHM)3: |
~3.1 nm w/25 m slit |
~4.6 nm w/25 m slit |
~7.6 nm w/25 m slit |
~ 9.5 nm w/25 m slit |
Signal-to-noise ratio at full signal4: |
>15000:1 @ 100 ms integration
>13000:1 @ 1000 ms integration |
>10000:1 @ 100 ms integration |
10000:1 @ 100 ms integration |
7500:1 @ 10 ms integration |
A/D resolution: |
16-bit |
16-bit |
16-bit |
16-bit |
Dark noise: |
6 RMS counts @ 100 ms
12 RMS counts @
1000 ms |
6 RMS counts @ 100 ms
12 RMS counts @
250 ms |
6 RMS counts @ 100 ms
12 RMS counts @ 250 ms |
8 RMS counts @ 10 ms
12 counts RMS @ 30 ms |
Dynamic range: |
15 x 10^6 (system); 5000:1 for a single acquisition |
15 x 10^6 (system); 1000:1 for a single acquisition |
5 x 10^6 (system); 5000:1 for a single acquisition |
3 x 10^5 (system); 5000:1 for a single acquisition |
Integration time5: |
1 ms-10 seconds |
1 ms-2000 ms |
1 ms-2000 ms |
1 ms-60 ms |
Corrected linearity: |
>99.8% |
>99.8% |
>99.8% |
>99.8% |
企业puter interfaces: |
USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud
|
Wireless (Wi-Fi) interface: |
Yes, with Remora adapter
|
Wi-Fi range: |
25 meters in free space
|
1 Useable range?is defined in the context of the NIRQuest models detector response and its typical grating response. For example, the 512-element detector has response at 850 nm, but grating response begins at 900 nm. The G9206 256-element detector response is sensitive to TEC temperature, and has response only to 2050 nm when the TEC is set to -20 C. The G9208 256-element detector has response to 2550 nm, but the grating efficiency drops off at 2500 nm.
2 Other filter options are available for order-sorting in the NIRQuest256-2.1 and NIRQuest256-2.5. Please contact an Applications Scientist for details.
3 Optical resolution (FWHM) depends on grating and slit selection.
4 SNR will decrease at longer integration times.
5 Maximum integration times are defined as the longest amount of time one can integrate the spectrometer before the dark level rises to half of full scale.
|